Page 19
5.
STERILIZATION PROGRAMS
5.1.
Program Parameters
Program
Icon
Name
Temp
Sterilization
time
(minutes)
Dry time
(minutes)
1.
Unwrapped
273.2°F
4
1 (default)
Range: 1-99
2.
Wrapped
273.2°F
7
20 (default)
Range: 20-99
3.
Unwrapped
pouches
273.2°F
7
15 (default)
Range: 1-99
4.
Wrapped pouches
273.2
7
30 (default)
Range: 20-30
5.
Unwrapped
249.8°F
20
1 (default)
Range: 1-99
6.
Wrapped
249.8°F
20
20 (default)
Range: 20-30
7.
Liquid A
249.8°F
15
0
8.
Liquid B -- waste
249.8°F
30
0
9.
Vacuum test
Vacuum test
10.
Bowie & Dick test
273.2
3.5
1