Connect count signal cable to CH0 to CH7 LEMO connector which are on front panel.

It' s possible to count maximum 8 channel signals.

 2-4-3.Start signal (TTL positive logic)

TTL positive logic signal starts counting action.

The width of start signal pulse must be more than 100ns.

It's possible to check the count start status by LED lamp on front panel.

It counts from start signal "ON" to stop signal "ON".

Without this signal you can alse start counting by software commands via LAN or USB.

 2-4-4.Stop signal (TTL positive logic)

TTL positive logic signal quits counting action.

The width of stop signal pulse must be more than 100ns.

It's possible to check the count stop status by LED lamp on front panel.

It counts from start signal "ON" to stop signal "ON".

Without this signal you can alse stop counting by software commands via LAN or USB.

 2-4-5.Gate signal (TTL positive logic)

Gate signal is TTL positive logic signal.

When gate signal goes "L" counting action stopped during "L" period.

When gate signal goes "L" LED lamp is "OFF" during this period.

LED lamp goes "ON" during gate signal is "H" when counter starts.

When gate signal is no connection, gate signal is equal to logic "H".

Run signal goes "H" when counting action is valid. It means count start and "Gate signal" are

in "H" logic, and LED goes "ON".

It's useful to use more than two counter units. Connecting this line to "Gate signal input" of another 

unit, another units go active counting action. Of course another units must be active in count mode 

before Run signal go "H".

Command characters are ASCII data.

The delimiter of communication command is CR+LF in transmit and receive case.

Type of command

ENCS

Enable to stop count by counter(CH7).

When count value of preset channel (CH7) reach to preset value, 

count action automatically stopped.

 2-4-2.Connect count signal

 2-4-6.Monitor out signal(In counting, TTL positive logic)

3.Explanation of communication commands

3-1.Communication commands

3-2.Read-out commands for counter set and the status of setting

 3-2-1.Stop enable by count value

8/15

Содержание NCT08-02

Страница 1: ...i Pre 300 0013 Japan Phone 81 0 29 832 3031 Fax 81 0 29 832 2662 URL http www tsujicon co jp APPLICATION OF ELECTRONIC DEVICES TSUJI ELECTRONICS CO LTD 3739 Kandatsu machi Tsuchiura city 8CH COUNTER T...

Страница 2: ...eans count on until STOP command enable counter stop enable to counter stop enable timer stop enable to timer stop mode read inquiry the mode read all counter and timer read all counter and timer set...

Страница 3: ...t and clear commands for counter and timer Read out and clear commands for all counter and timer Read out and clear commands for all counter Read out and clear commands for timer Communication command...

Страница 4: ...del of NCT08 01 The commands are almost the same of the NCT08 01 The preset value is round down to multiple number of eight It shows the beginning of count action LED ON means the counter gate open In...

Страница 5: ...vel for each channel Please confirm the signals are between the permitted level 12mA 36mA 1 4mA 20mA 0 The input impedance is 50 so the voltage level will be 0 6V 1 8V 1 and 0 2V 0 To control this one...

Страница 6: ...m 192 168 1 55 bytes 32 time 1ms TTL 255 If replies are listed above physical connection is correct If physical connection is not correct communication replies are these In this case please retry the...

Страница 7: ...play Then select 1 push return push return push return enter port address of then push return Default port NO is if you need to change this number recommended port number is 10000 to 10999 push return...

Страница 8: ...ections specify the folders of downloaded driver software When the driver software installs are done open the Control Panel START Control Panel click the System icon Then property display of system ap...

Страница 9: ...nal When gate signal goes L counting action stopped during L period When gate signal goes L LED lamp is OFF during this period LED lamp goes ON during gate signal is H when counter starts When gate si...

Страница 10: ...Setting and readout of preset counter Type of command dddd To set preset value dddd in decimal to preset counter Unit is Kcts Maximum value is 281 474 976 710 Kcts dddd To set preset value dddd in de...

Страница 11: ...Reply 123456 Reply data is 8 digit decimal value 4294967000 If reply data is over 8 digit the digit is as the timer digit Unit is s Type of command STRT Enable counter start Type of command ST Enable...

Страница 12: ...is as the value digit Type of command To clear timer value Inquiry of over flow count data Type of command To inquiry about data overflow timer and count data Reply example over Timer is in normal ov...

Страница 13: ...er number 4 overflow Reply data 04 mean Counter number 6 overflow Reply for command FLG 2 Time up carry monitor RUN OUT COUNTER ON Timer overflow Counter number 7 overflow TTL GATE TTL STOP TTL START...

Страница 14: ...ower line of NCT08 02 Turn on power of NCT08 02 again the firmware version will be new Without turning off power line it gets start by RESET command There are two pieces of flash ROM FROM loaded insid...

Страница 15: ...a 100 times total disable period is 120ns 100 12 s Disable period for each reading access is always same 120ns Counter and timer stops simultaneously you d better to read data by minimum command When...

Страница 16: ...p counting action continues until stop trigger or STOP command Enable to count or not for all channels and timer Goes to count mode by open or H level In count mode LED turn on green In count mode sig...

Отзывы: