 0 Server configuration

 1 Channel 1 configuration

 3 E-mail settings

 5 Expert settings

 6 Security

 7 Factory defaults

 8 Exit without save

 9 Save and exit           Your choise ?

In these setting items, only IP address must be set. Other items may be no change to set.

If you enter wrong number by mistake, you'd better to retore above data.

 2-2-3.Turn back PC set data

If you change setting data of PC, turn back PC data to initial condition.

Let's try connecting test by optional Windows software such as "telnet".

Example test data is "telnet 192.168.1.55 7777" in MS-DOS prompt screen.

(IP address number must to be set number preliminary. )

When in telnet operation, send data "VER?" which are expected reply data.

If reply data is like ”1.00 05-07-07", then communication line is right.

"Telnet" function contains hardware and software function, when disconnect line, line must be cut off 

by "telnet" software operation before hardware disconnection.

 Driver software for USB (Universal Serial Bus) operation is downloaded from our web server.

By installing this software, connecting is done like a COM port connection.

Switch on a NCT08-02, connect USB cable, then there will be message box that shows "New hardware 

device was found". According to PC directions, specify the folders of downloaded driver software. 

 When the driver software installs are done, open the "Control Panel" ( START → Control Panel), 

click the "System" icon. Then property display of system appears. Select the "Device Manager". 

Confirm the USB Serial Port (Com*) number in the tree between Port display.

 Select the USB Serial Port (Com*), click Advanced in Port Setting display. It's possible to change

COM Port number. Click the check box Disable PNP □ , it's possible to short operation time in

connecting USB cable again.

 Launch RS-232C communication software such as hyper-terminal included in your PC.

Set COM port number which is known in previous item, it's need to check communication line is available

or not. In this case It's no need to set baud rate. Send the command data such as "VER?", then get  

the reply  "1.00 04-09-03", there is no problem for communication 

Connect the communication cable to USB port or LAN port on rear panel.

There is no need to connect both side, but there will be no problem even if connect both side.

If connected both side, latest command will be valid to the unit.

 2-2-4.Connecting test

2-3.Setting USB communication

 2-3-1.Preparations for setting

 2-3-2.Connecting test

2-4.Connect signal cables

 2-4-1.Connect communication cables

7/15

Содержание NCT08-02

Страница 1: ...i Pre 300 0013 Japan Phone 81 0 29 832 3031 Fax 81 0 29 832 2662 URL http www tsujicon co jp APPLICATION OF ELECTRONIC DEVICES TSUJI ELECTRONICS CO LTD 3739 Kandatsu machi Tsuchiura city 8CH COUNTER T...

Страница 2: ...eans count on until STOP command enable counter stop enable to counter stop enable timer stop enable to timer stop mode read inquiry the mode read all counter and timer read all counter and timer set...

Страница 3: ...t and clear commands for counter and timer Read out and clear commands for all counter and timer Read out and clear commands for all counter Read out and clear commands for timer Communication command...

Страница 4: ...del of NCT08 01 The commands are almost the same of the NCT08 01 The preset value is round down to multiple number of eight It shows the beginning of count action LED ON means the counter gate open In...

Страница 5: ...vel for each channel Please confirm the signals are between the permitted level 12mA 36mA 1 4mA 20mA 0 The input impedance is 50 so the voltage level will be 0 6V 1 8V 1 and 0 2V 0 To control this one...

Страница 6: ...m 192 168 1 55 bytes 32 time 1ms TTL 255 If replies are listed above physical connection is correct If physical connection is not correct communication replies are these In this case please retry the...

Страница 7: ...play Then select 1 push return push return push return enter port address of then push return Default port NO is if you need to change this number recommended port number is 10000 to 10999 push return...

Страница 8: ...ections specify the folders of downloaded driver software When the driver software installs are done open the Control Panel START Control Panel click the System icon Then property display of system ap...

Страница 9: ...nal When gate signal goes L counting action stopped during L period When gate signal goes L LED lamp is OFF during this period LED lamp goes ON during gate signal is H when counter starts When gate si...

Страница 10: ...Setting and readout of preset counter Type of command dddd To set preset value dddd in decimal to preset counter Unit is Kcts Maximum value is 281 474 976 710 Kcts dddd To set preset value dddd in de...

Страница 11: ...Reply 123456 Reply data is 8 digit decimal value 4294967000 If reply data is over 8 digit the digit is as the timer digit Unit is s Type of command STRT Enable counter start Type of command ST Enable...

Страница 12: ...is as the value digit Type of command To clear timer value Inquiry of over flow count data Type of command To inquiry about data overflow timer and count data Reply example over Timer is in normal ov...

Страница 13: ...er number 4 overflow Reply data 04 mean Counter number 6 overflow Reply for command FLG 2 Time up carry monitor RUN OUT COUNTER ON Timer overflow Counter number 7 overflow TTL GATE TTL STOP TTL START...

Страница 14: ...ower line of NCT08 02 Turn on power of NCT08 02 again the firmware version will be new Without turning off power line it gets start by RESET command There are two pieces of flash ROM FROM loaded insid...

Страница 15: ...a 100 times total disable period is 120ns 100 12 s Disable period for each reading access is always same 120ns Counter and timer stops simultaneously you d better to read data by minimum command When...

Страница 16: ...p counting action continues until stop trigger or STOP command Enable to count or not for all channels and timer Goes to count mode by open or H level In count mode LED turn on green In count mode sig...