background image

Power

NIM AC85V~132V 0.2A, DC +6V:0.4A,  -6V:0.3A

Counter

Input level

 Current Level -12mA~-36mA:"1"  -4mA~+20mA:"0"

Input

NIM

  Zin = 50Ω

  

 Count frequency More than 500MHz

Input connector  LEMO ERA00250
Channels

 0~7channels (CH7 is preset counter)

Overflow digit

 When count value over the limit send the data "over"

Digit of count

 48bit data(0~281,474,976,710,655 cts)

Timer

 1channel 40bit 1~1,099,511,627,775 μs

Resolution

 0.000001s (1μs)   

Accuracy
Preset time

 1 ~ 1,099,511,627,775 μs

Special

  1channel CH7 Fixed

counter

Preset count

  1 ~ 281,474,976,710,655 cts

Counter

single mode

Count starts by start trigger or "START" command

Mode

Enable to count within preset value one time
Temporarily turn into stop count by stop trigger or "STOP"

command

When disable to preset counter stop and preset timer stop

counting action continues until stop trigger or "STOP"
command

Enable to count or not for all channels and timer
Goes to  count mode by open or "H" level
In count mode LED turn on "green"
In count mode signal out goes "H"
Connect this port to "Gate in TTL" of next unit.
It's possible to activate multiple units simultaneously
Count start input(by TTL rising edge), count stop input
        (by TTL rising edge)
LAN, USB
Enable to upgrade firmware software by communication
NIM 1 UNIT

For the further information, feel free to ask us.

8.Performance and specifications

  0.005%

Gate in TTL

LED in counting
Signal out
In counting

Control input

Number of channel

Number of channel

3739, Kandatsu-machi, Tsuchiura-city, Ibaraki 300-0013, Japan

 TEL: +81-(0)29-832-3031   FAX: +81-(0)29-832-2662

 E-mail : [email protected]

URL : http://www.tsujicon.jp

Communication
Version upgrade
Case

Tsuji-Electronics Co.,Ltd

CH0~CH8
COUNT IN

GND

LEMO

GND

51

NIM LEVEL

NIM LEVEL
RECEIVER

ECL
COUNTER

15/15

Содержание NCT08-02

Страница 1: ...i Pre 300 0013 Japan Phone 81 0 29 832 3031 Fax 81 0 29 832 2662 URL http www tsujicon co jp APPLICATION OF ELECTRONIC DEVICES TSUJI ELECTRONICS CO LTD 3739 Kandatsu machi Tsuchiura city 8CH COUNTER T...

Страница 2: ...eans count on until STOP command enable counter stop enable to counter stop enable timer stop enable to timer stop mode read inquiry the mode read all counter and timer read all counter and timer set...

Страница 3: ...t and clear commands for counter and timer Read out and clear commands for all counter and timer Read out and clear commands for all counter Read out and clear commands for timer Communication command...

Страница 4: ...del of NCT08 01 The commands are almost the same of the NCT08 01 The preset value is round down to multiple number of eight It shows the beginning of count action LED ON means the counter gate open In...

Страница 5: ...vel for each channel Please confirm the signals are between the permitted level 12mA 36mA 1 4mA 20mA 0 The input impedance is 50 so the voltage level will be 0 6V 1 8V 1 and 0 2V 0 To control this one...

Страница 6: ...m 192 168 1 55 bytes 32 time 1ms TTL 255 If replies are listed above physical connection is correct If physical connection is not correct communication replies are these In this case please retry the...

Страница 7: ...play Then select 1 push return push return push return enter port address of then push return Default port NO is if you need to change this number recommended port number is 10000 to 10999 push return...

Страница 8: ...ections specify the folders of downloaded driver software When the driver software installs are done open the Control Panel START Control Panel click the System icon Then property display of system ap...

Страница 9: ...nal When gate signal goes L counting action stopped during L period When gate signal goes L LED lamp is OFF during this period LED lamp goes ON during gate signal is H when counter starts When gate si...

Страница 10: ...Setting and readout of preset counter Type of command dddd To set preset value dddd in decimal to preset counter Unit is Kcts Maximum value is 281 474 976 710 Kcts dddd To set preset value dddd in de...

Страница 11: ...Reply 123456 Reply data is 8 digit decimal value 4294967000 If reply data is over 8 digit the digit is as the timer digit Unit is s Type of command STRT Enable counter start Type of command ST Enable...

Страница 12: ...is as the value digit Type of command To clear timer value Inquiry of over flow count data Type of command To inquiry about data overflow timer and count data Reply example over Timer is in normal ov...

Страница 13: ...er number 4 overflow Reply data 04 mean Counter number 6 overflow Reply for command FLG 2 Time up carry monitor RUN OUT COUNTER ON Timer overflow Counter number 7 overflow TTL GATE TTL STOP TTL START...

Страница 14: ...ower line of NCT08 02 Turn on power of NCT08 02 again the firmware version will be new Without turning off power line it gets start by RESET command There are two pieces of flash ROM FROM loaded insid...

Страница 15: ...a 100 times total disable period is 120ns 100 12 s Disable period for each reading access is always same 120ns Counter and timer stops simultaneously you d better to read data by minimum command When...

Страница 16: ...p counting action continues until stop trigger or STOP command Enable to count or not for all channels and timer Goes to count mode by open or H level In count mode LED turn on green In count mode sig...

Отзывы: