Type of command
CTR?xx
To read count value of channel number xx.
CTR?xxyy
To read count value from channel number xx to channel number yy.
Reply
1234567890
Reply data is 10 digit data in decimal. Unit is cts.
1234567890 2345678901 3456789012 ・・・・・ 0123456789
Reply data is 10 digit data in decimal in order. Unit is cts.
If reply data is over 10 digit, the digit is as the value digit.
Type of command
CLCTxx
To clear the count value of counter channel xx(00~07).
CLCTxxyy
To clear the count value from counter channel xx to channel yy.
CLPC
To clear the count value of preset counter channel 07.
Type of command
TMR?
To read timer value.
Reply
1234567890
Reply data is 10 digit data in decimal. Unit is μs.
If reply data is over 10 digit, the digit is as the value digit.
Type of command
CLTM
To clear timer value.
3-5.Inquiry of over-flow count data
Type of command
ALM?
To inquiry about data overflow timer and count data
Reply example
overXXXX--
Timer is in normal.
overXXXXTM
Timer is overflow.
XXXX means overflow counter channel. XXXX is 4 digit hex data.
(Example)
over0001--
It means counter CH0 is overflow.
[0000 0001]
over0009--
It means counter CH0 and CH3 are overflow.
[0000 1001]
over0039--
It means counter CH0,CH3, CH4 and CH5 are overflow.
[0011 1001]
over000ATM
It means counter CH1,CH3 and timer are overflow.
[0000 1010]
over0000--
It means no overflow.
[0000 0000]
Type of command
VER?
To inquiry about software version information
Reply example
1.00 07-01-05 NCT08-02
Reply is in order, version number, date and type of the unit.
3-4-2.Read-out and clear commands for all counter
3-4-3.Read-out and clear commands for timer
3-6.Inquiry of software version information
11/15