background image

1.Specifications

1-1.Abstract

 This unit has fast eight channel 48 bit counters and a 40 bit timer.

It counts during timer set interval or until the limited set count value.

Setting interval for count is 1μs to 1,099,511.627775s. Setting limited count value is 1cts 

to 281,474,976,710,655cts. 

*1)

This unit is assembled in a NIM 1 unit case. By connecting several units by RUN-GATE connection, 

total count channels can be increased to 8×n channels.

It's possible to control by LAN or USB communication.

The NCT08-02 is made as an upgrade model of NCT08-01. The commands are almost the same of the

NCT08-01.

*) The preset value is round down to multiple number of eight.

     ③

 ①

 ②

 ④

 ⑤

  ⑥

 ⑦

It shows the beginning of count action. LED "ON" means the counter gate open.

Input count connectors. There are CH0 to CH7. (INPUT)

  It's possible to set preset value only for CH7.

Connect TTL level signal (3.3 - 5V) from external unit.

START:

When upward edge signal is "ON" ("H"), it starts counting. (INPUT)

Normally (not connected) START signal is "L".

When count mode is selected to "Counter 07 stop" or  "Timer stop" and count

 value or timer value goes on limit, START can't become true.

NCT08-02 User's manual

1-2.Appearance

COUNTER/TIMER
      NCT08-02

LAN

COUNT

GATE

START

STOP

TTL IN

RUN

TTL OUT

CH0

CH1

CH2

CH3

CH4

CH5

CH6

CH7

USB

3/15

Содержание NCT08-02

Страница 1: ...i Pre 300 0013 Japan Phone 81 0 29 832 3031 Fax 81 0 29 832 2662 URL http www tsujicon co jp APPLICATION OF ELECTRONIC DEVICES TSUJI ELECTRONICS CO LTD 3739 Kandatsu machi Tsuchiura city 8CH COUNTER T...

Страница 2: ...eans count on until STOP command enable counter stop enable to counter stop enable timer stop enable to timer stop mode read inquiry the mode read all counter and timer read all counter and timer set...

Страница 3: ...t and clear commands for counter and timer Read out and clear commands for all counter and timer Read out and clear commands for all counter Read out and clear commands for timer Communication command...

Страница 4: ...del of NCT08 01 The commands are almost the same of the NCT08 01 The preset value is round down to multiple number of eight It shows the beginning of count action LED ON means the counter gate open In...

Страница 5: ...vel for each channel Please confirm the signals are between the permitted level 12mA 36mA 1 4mA 20mA 0 The input impedance is 50 so the voltage level will be 0 6V 1 8V 1 and 0 2V 0 To control this one...

Страница 6: ...m 192 168 1 55 bytes 32 time 1ms TTL 255 If replies are listed above physical connection is correct If physical connection is not correct communication replies are these In this case please retry the...

Страница 7: ...play Then select 1 push return push return push return enter port address of then push return Default port NO is if you need to change this number recommended port number is 10000 to 10999 push return...

Страница 8: ...ections specify the folders of downloaded driver software When the driver software installs are done open the Control Panel START Control Panel click the System icon Then property display of system ap...

Страница 9: ...nal When gate signal goes L counting action stopped during L period When gate signal goes L LED lamp is OFF during this period LED lamp goes ON during gate signal is H when counter starts When gate si...

Страница 10: ...Setting and readout of preset counter Type of command dddd To set preset value dddd in decimal to preset counter Unit is Kcts Maximum value is 281 474 976 710 Kcts dddd To set preset value dddd in de...

Страница 11: ...Reply 123456 Reply data is 8 digit decimal value 4294967000 If reply data is over 8 digit the digit is as the timer digit Unit is s Type of command STRT Enable counter start Type of command ST Enable...

Страница 12: ...is as the value digit Type of command To clear timer value Inquiry of over flow count data Type of command To inquiry about data overflow timer and count data Reply example over Timer is in normal ov...

Страница 13: ...er number 4 overflow Reply data 04 mean Counter number 6 overflow Reply for command FLG 2 Time up carry monitor RUN OUT COUNTER ON Timer overflow Counter number 7 overflow TTL GATE TTL STOP TTL START...

Страница 14: ...ower line of NCT08 02 Turn on power of NCT08 02 again the firmware version will be new Without turning off power line it gets start by RESET command There are two pieces of flash ROM FROM loaded insid...

Страница 15: ...a 100 times total disable period is 120ns 100 12 s Disable period for each reading access is always same 120ns Counter and timer stops simultaneously you d better to read data by minimum command When...

Страница 16: ...p counting action continues until stop trigger or STOP command Enable to count or not for all channels and timer Goes to count mode by open or H level In count mode LED turn on green In count mode sig...

Отзывы: