The set value of "density line slope" of the
arithmetic expression for calculating the density
from the phase measurement data, etc. can be
verified.
The set value of "density intercept" of the arithmetic
expression for calculating the density from the
phase measurement d ata, etc. can be verified.
Normally, this is set to zero.
Upon getting into [2: SETTING MENU], the current
output and the LED density display are switched to
simulated values. In this screen, the set value of the
simulated value (unit: %TS) can be verified.
In external synchronized operations, the set value
of the delayed time (unit: minute) from when the
external contact input is turned ON until the
measurement starts can be verified. For details,
refer to Section 6.7.
The phase at the time of zero calibration can be
verified.
[ESC]
(Previous
Menu)
[ESC]
(Previous
Menu)
[ESC]
(Previous
Menu)
[ESC]
(Previous
Menu)
[ESC]
(Previous
Menu)
a
:
DENSITY LINE SLOPE
DATA
:
0.0840
[ESC] RETURN
b
:
DENSITY INTERCEPT
DATA
:
0.0
[ESC] RETURN
ot
:
DENSTY TEST OUT
DATA
:
1.5
[ESC] RETURN
dt
:
DELAYED SYNC.TIME
DATA
:
0.5
[ESC] RETURN
zp:ZERO POINT PHASE
DATA
:
80.00
[ESC] RETURN
[DN]
(Pre. menu)
[UP]
(Next menu)
[DN]
(Pre. menu)
[DN]
(Pre. menu)
[UP]
(Next menu)
[DN]
(Pre. menu)
[UP]
(Next menu)
[DN]
(Pre. menu)
[UP]
(Next menu)
[UP]
(Next menu)
The fluid temperature at the time of zero calibration can
be verified.
[ESC]
(Previous
Menu)
[DN]
(Pre. menu)
[UP]
(Next menu)
zT:ZERO WATER TEMP.
DATA
:
25.00
[ESC] RETURN
Содержание Density (Consistency) Meter LQ500
Страница 1: ...6 F 8 A 0 5 2 1 OPERATION MANUAL DENSITY METER TYPE LQ500 ...
Страница 2: ......
Страница 51: ... 49 6 F 8 A 0 5 2 1 ...
Страница 61: ... 59 6 F 8 A 0 5 2 1 ...
Страница 93: ... 91 6 F 8 A 0 5 2 1 mode from the setting mode to resume the normal measurement ...
Страница 113: ......