74
6
F
8
A
0
5
2
1
Table. 8.2
Self-diagnosis data
Data Item
Symbol
Item name
Data Status
Data (Numeric value range;
normal data range, status)
ST
Status
[STATUS]
•
Normal [GOOD]
•
Warning [WARNING.]
Whether the phase
measurement operation is
normally functioning or not can
be verified.
SL
Microwave
recieved signal
level
[MICROWAVE
SIG. LVL]
Without error judgment based
on data numeric
•
Usual range: -90 to -40 (dBm)
F
Microwave coef.
[
MICROWAV
E COEF.
]
Without error judgment based
on data numeric
•
Normal range: 1825
〜
1975
G
Data
concerning
microwave
phase
measurement
[RF DATA]
Without error judgment based
on data numeric
•
Normal range + 10.0
〜
+ 80.0 (
°
C)
J
+5 V voltage
[+5V POWER
SUPPLY]
Without error judgment based
on data numeric
•
Normal range
4.5
〜
5.5
(
V
)
pd
Reference
phase error
[
REF
PHASE
]
Without error judgment based
on data numeric
•
Normal range
- 9.9
〜
+ 9.9 (
°
C)
P
Memory check
[
MEMORY
CHECK
]
Normal
[
GOOD
]
Abnormal
[N.G.]
EPROM, RAM, EEPROM are
checked