Fig. 3.5.1 External connection
Fig. 3.6 Terminals inside the converter
Ground (PE)
(Ground resistance: 100
Ω
or less)
Density meter error or
maintenance-in-progress contact
output
Grounding terminal
Note: The ground should be connected to either the PE
terminal of the terminal bock inside converter or to the
grounding terminal of the case.
Density correction multiplier switching
voltage input signal
(H: 20-30VDC, L: 2VDC or less, input
resistance about 3k
Ω
)
External sync contact input
(Contact capacity: 24VDC, 1A or more)
Measured density output
(4-20mADC, 750
Ω
or less, isolated
CVV
CVVS
Converter
Power 100 to 240VAC, 50/60Hz
CVV
DI
COM
COM1
DO1
FG
AO+
AO-
COM2
DI2
L1
L2
DI3
Tx
+
Tx-
Rx
+
Rx-
SG
FG
+24
V
0V
FG
Tx
+
Tx-
Rx+
Rx-
SG
FG
+24
V
0V
Power
cable
Communication
cable
RF
section
FG
Ec-
Ec
+
Conductivity
signal input
(4 ~ 20mADC)
Note: The FG wires of the dedicated
cables A and B should not be
connected to the FG terminal on the
detector (measuring section) side.
FG
CVVS
CVVS
Ground
(Ground resistance: 100
Ω
or less)
Power supply
Terminal block
+24V
0V
T x
+
Rx+
NC
(FG)
DI
D01
DI2
DI3
AO+
L1
L2
FG
T x
−
Rx
−
SG
(FG)
COM
COM1
(FG)
COM2
AO
−
Communication
External sync input DI,
COM
Density meter error or
maintenance-in-progress
signal: DO1, COM1
Multiplier switching:
DI2, DI3, COM2
AO (+,
−
)
4-20mA
Density output
(HART)
AC Power
Содержание Density (Consistency) Meter LQ500