10.3.4 Setting the change-rate limit
(1) Switching to the setting mode (see Subsection 5.2.5)
First of all, press the [ESC] key of the converter several times (normally once although this
varies with the operation status) to return to the initial menu display. Next, use the [
→
] key to
move the LCD indicator cursor to the menu number "2" of "2 SETTING MENU" and then
press the [SET] key to display the warning message saying "Test output will be valid". Make
sure that there is no problem and then press [
→
] to get into the setting mode. And return to [2:
SETTING MENU]. Then, the output will be switched to the simulated output that is set
beforehand.
(2) Selecting the parameter setting menu (see Subsection 5.2.5)
In the state that the menu list of menu numbers 5 to 8 is displayed, use the [
→
] key to move the
LCD indicator cursor to the menu number "5" of "5: SET PARAMETERS" and then press the
[SET] key to select the menu of "5: SET PARAMETERS".
(3) Verifying or changing the change-rate limit width (see Subsection 5.2.16)
Press the [DN] key two times (or the [UP] key twelve times) to display "dx: CHANGE
RATE LIMIT". On this screen, it is possible to verify the set value of the change-rate limit
width. If it is necessary to change the setting, press the [SET] key here to switch to the setup
display of the change-rate limit width and enter an appropriate value for the width. Use the
[
→
] key to move to the digit, and use the [UP] or [DN] key to alter the numeric value of the
relevant digit. When entering input is completed, press the [SET] key to confirm it. The screen
will be returned to the immediately preceding one; make sure that the value is correctly reset.
(4) Verifying or changing the times of change-rate limit (see Subsection 5.2.16)
Continuing on from the operation in (3) above, press the [UP] key once to display "HL: LIMIT
TIMES". On this screen, it is possible to verify the set value of the change-rate limit times. If
it is necessary to change the setting, press the [SET] key here to switch to the setup display of
the change-rate limit times and enter an appropriate value for the times. Use the [
→
] key to
move to the digit, and use the [UP] or [DN] key to alter the numeric value of the relevant digit.
When entering input is completed, press the [SET] key to confirm it. The screen will be
returned to the immediately preceding one; make sure that the value is correctly reset.
Note: If the times of change-rate limit is set to zero, the change-rate limit function is
made inactive.
(5) Returning to the measuring mode
Press the [ESC] key twice to return to the initial menu screen and return to the measuring
mode from the setting mode to resume the normal measurement.
Содержание Density (Consistency) Meter LQ500
Страница 1: ...6 F 8 A 0 5 2 1 OPERATION MANUAL DENSITY METER TYPE LQ500 ...
Страница 2: ......
Страница 51: ... 49 6 F 8 A 0 5 2 1 ...
Страница 61: ... 59 6 F 8 A 0 5 2 1 ...
Страница 93: ... 91 6 F 8 A 0 5 2 1 mode from the setting mode to resume the normal measurement ...
Страница 113: ......