(4) Setting density multipliers C2, C3 and C4
Continuing on from (3), press the [UP] key once to switch to the display of "C2:
DensityMultiplier". Switch to the setup display by pressing the [SET] key here and then input
the density multiplier C2. Use the [
→
] key to move to the digit, and use the [UP] or [DN] key
to alter the numeric value of the relevant digit. When entering input is completed, press the
[SET] key to confirm it. The screen will be returned to the immediately preceding one; make
sure that the value is correctly reset. Continue on to switch to the next setup menu display with
the [UP] key and set C3 and C4 in accordance with the same procedure. When setting the
density multiplier C1, refer to "6.5 Span Calibration".
(5) Returning to the measuring mode
Press the [ESC] key twice to return to the initial menu screen and return to the measuring
mode from the setting mode to resume the normal measurement.