5.2.11 Zero calibration display and operating procedures
5.2.12 Span calibration display and operating procedures
Move the cursor to "7" with [
→
] key, and press [SET] key
to select [7:SPAN CALIBRATION]
5:SET PARAMETERS
6:ZERO CALIBRATION
7:SPAN CALIBRATION
8:ANGLE ROTATION
C1:DensityMultiplier
DATA:1.000
[SET] CHANGE
[ESC] RETURN
[ESC]
(Previous
Menu)
C1:DensityMultiplier
RANGE :0.000-9.999
DATA : 1.000
[SET] SET,[ESC]CANCEL
[SET](Set)
[SET](Fix)
[ESC](Cancel)
The density multiplier C1 should be set to the suitable
value so that the measured density value is
corresponding to the manual analysis value. For
details, refer to Section 6.5.
Move the cursor to "6" with [
→
] key, and press [SET] key to select [6:ZERO CALIBRATION]
5:SET PARAMETERS
6:ZERO CALIBRATION
7:SPAN CALIBRATION
8:ANGLE ROTATION
ZERO CAL
*
p PHASE:
87.93
°
N=0
T TEMP.:
26.55
°
C
X DENSITY: 0.02 %TS
ZERO POINT
CALIBRATION
HAS BEEN COMPLETED
[SET](Set)
New data displayed
[ESC]
(Previous
Menu)
Press the [SET] key for zero calibration. In zero calibration, the
phase and the fluid temperature of zero point will be replaced
with the present measured values. And also, angle rotation "N"
will be set to zero.
Note:
The display will return to the left display automatically
after approximately one second.
Содержание Density (Consistency) Meter LQ500
Страница 1: ...6 F 8 A 0 5 2 1 OPERATION MANUAL DENSITY METER TYPE LQ500 ...
Страница 2: ......
Страница 51: ... 49 6 F 8 A 0 5 2 1 ...
Страница 61: ... 59 6 F 8 A 0 5 2 1 ...
Страница 93: ... 91 6 F 8 A 0 5 2 1 mode from the setting mode to resume the normal measurement ...
Страница 113: ......