38
6
F
8
A
0
5
2
1
5.2.6
Measuring mode display and operating procedures
5.2.7
Reading of parameters display and operating procedures
OP
:
MEASURING MODE
DATA
:
CONT
[SET] CHANGE
[ESC] RETURN
OP:MEASURING MODE
RANGE :CONT/EXT
DATA :CONT
[SET] SET,[ESC]CANCEL
1
:
MONITORING MENU
2
:
SETTING MENU
3
:
MEASURING MODE
Move the cursor to "3" with [
→
] key, and press [SET] key.
Press the SET key to set.
Press the SET key to confirm.
Press the ESC key to cancel
.
Each time the
゚
[UP] or [DN] key is pressed, CONT/EXT are
mutually alternated thus making it possible to select an operation
mode. Select "CONT" for normal continuous operations; select
"EXT" for external synchronized operations. For details on the
external synchronized operation, refer to Section 6.7.
[ESC]
(Previous
Menu)
Move the cursor to "1" with [
→
] key, and press [SET] key,
and select [1 : READ PARAMETERS]
UR:UPPER RANGE
DATA
:
3.0
[ESC] RETURN
LR:LOWER RANGE
DATA
:
0.0
[ESC] RETURN
1
:
READ PARAMETERS
2
:
MEASURED VALUES
3
:
SELF-DIAGNOSIS
C
:
DENSITY MULTIPLIER
DATA
:
1.000(C1)
[ESC] RETURN
The set value of the density multiplier C, which is used for
density calculation, can be verified. If C2, C3 or C4 is
displayed in the parentheses, it indicates that the density
multiplier switched to by the external voltage signal (DI) is
selected.
The set value of the upper density measurement range
(the density whose current output is 20mA) can be verified.
The set value of the lower density measurement range
(the density whose current output is 4mA) can be verified.
[ESC]
(Previous
Menu)
[ESC]
(Previous
Menu)
[ESC]
(Previous
Menu)
[DN]
(Pre. menu)
[UP]
(Next menu)
[DN]
(Pre. menu)
[UP]
(Next menu)
[UP]
(Next menu)
Содержание Density (Consistency) Meter LQ500
Страница 1: ...6 F 8 A 0 5 2 1 OPERATION MANUAL DENSITY METER TYPE LQ500 ...
Страница 2: ......
Страница 51: ... 49 6 F 8 A 0 5 2 1 ...
Страница 61: ... 59 6 F 8 A 0 5 2 1 ...
Страница 93: ... 91 6 F 8 A 0 5 2 1 mode from the setting mode to resume the normal measurement ...
Страница 113: ......