42
6
F
8
A
0
5
2
1
[ESC]
(Previous
Menu)
The micro wave signal constant can be verified.
Normally, the value is 1825 to 1975.
[ESC]
(Previous
Menu)
[ESC]
(Previous
Menu)
[ESC]
(Previous
Menu)
[ESC]
(Previous
Menu)
[ESC]
(Previous
Menu)
The soundness of the phase measurement unit
can be verified.
The soundness of various memory units can be
verified.
The converter's RF data (concerning the micro
wave phase measurement) can be verified.
Normally, the value is 10 to 70.
The voltage of the 5-volt power supply can be
checked.
G:RF DATA.
DATA : 50.00
[ESC] RETURN
J:+5V POWER SUPPLY
DATA :5.00
[ESC] RETURN
Mc:MEMORY CHECK:GOOD
EPROM :GOOD
RAM :GOOD
EEPROM :GOOD
ST:STATUS :GOOD
[ESC] : RETURN
pd:REF PHASE
DATA
:
0.0
[ESC] RETURN
F:MICROWAVE COEF.
DATA : 1910
[ESC] RETURN
[DN]
(Pre. menu)
[UP]
(Next menu)
[DN]
(Pre. menu)
[UP]
(Next menu)
[DN]
(Pre. menu)
[UP]
(Next menu)
[DN]
(Pre. menu)
[UP]
(Next menu)
[DN]
(Pre. menu)
[UP]
(Next menu)
[DN]
(Pre. menu)
Содержание Density (Consistency) Meter LQ500
Страница 1: ...6 F 8 A 0 5 2 1 OPERATION MANUAL DENSITY METER TYPE LQ500 ...
Страница 2: ......
Страница 51: ... 49 6 F 8 A 0 5 2 1 ...
Страница 61: ... 59 6 F 8 A 0 5 2 1 ...
Страница 93: ... 91 6 F 8 A 0 5 2 1 mode from the setting mode to resume the normal measurement ...
Страница 113: ......