5.2 Menus and operations
Operations should be done with five keys for setting, in combination with the LCD display.
This section shows menus and operations.
5.2.1.
Main menu
Main menu is composed of three basic menus shown below. Table 5.2.1 shows the functions of
each menu and performances when selected.
<
main menu>
Table 5.2.1 Functions and performances of main menu
1
:
MONITORING MENU
2
:
SETTING MENU
3
:
MEASURING MODE
Functions
Reading of each
measuring conditions
(parameters), measured
values, and self-diagnosis
data
Changing of each
measuring conditions
(parameters), zero
calibration and span
calibration
Mode selection from among
two measuring modes
(operation modes) of the
normal continuous operation
and the externally
synchronized operation
Measured density
Output (4 ~ 20mA)
Measured density
continuous output
Density Test output
Measured density continuous
output
LCD Density
display
Measured density value
Density Test output
Measured density valve
[Measure] indicator
On
Off
On
Note: “Measured density value is output instead of “Density Test output” as the LCD density display
on the panel when “Zero calibration” or “Span calibration” is selected in the setting menu. This
arrangement is intended to compare the measured density values before and after the
calibration for both Zero and Span calibrations. As to the measured density output (4
−
20 mA),
“Density Test output” is used for all menu items including Zero calibration and Span calibration.
1 :
MONITORING MENU
2 :
SETTING MENU
3 :
MEASURING MODE
Содержание Density (Consistency) Meter LQ500
Страница 1: ...6 F 8 A 0 5 2 1 OPERATION MANUAL DENSITY METER TYPE LQ500 ...
Страница 2: ......
Страница 51: ... 49 6 F 8 A 0 5 2 1 ...
Страница 61: ... 59 6 F 8 A 0 5 2 1 ...
Страница 93: ... 91 6 F 8 A 0 5 2 1 mode from the setting mode to resume the normal measurement ...
Страница 113: ......