Error category
Priority
Error
number
Indication in the
display
Description
70
II
702
System failure
Operating system error, please
restart system. In the case of
failure contact your responsible
TNI medical AG representative.
70
II
703
System failure
EEPROM defective, please
restart system. In the case of
failure contact your responsible
TNI medical AG representative
70
II
704
System failure
User settings damaged, please
restart system. In the case of
failure contact your responsible
TNI medical AG representative.
70
II
705
System failure
Firmware not usable with this
device, please restart system. In
the case of failure contact your
responsible TNI medical AG
representative.
70
I
751
SD-card not avail-
able
SD card not available
70
I
752
SD-card or file
defective
SD-card checksum error
70
I
753
System failure
Battery voltage too low, please
restart system. In the case of
failure contact your responsible
TNI medical AG representative.
70
I
754
System failure
Firmware checksum error (on
SD card update), please restart
system. In the case of failure
contact your responsible TNI
medical AG representative.
70
I
755
System failure
Firmware not usable with this
device (on SD card update),
please restart system. In the
case of failure contact your
responsible TNI medical AG
representative.
80
I
851
Clean the filter
NOTE: change filter; filter
should be changed every 3
month
Tabelle 5 - Error codes