Error category
Priority
Error
number
Indication in the
display
Description
Motor tempera-
ture is higher
than 120 °C.
II
403
Blower sensor
defective
Blower temperature sensor
defective
40
II
404
Fan blower defec-
tive
Fan blower defective, please
contact your responsible TNI
medical AG representative
50
II
501
-
Display defect, please restart
system. In the case of failure
contact your responsible TNI
medical AG representative.
Display defect. Error is not dis-
played, but you should hear an
alarm signal.
50
II
502
System failure
Sensor errors on system
startup, please restart system.
In the case of failure contact
your responsible TNI medical
AG representative.
60
II
601
Sensor defective
Temperature sensor applicator
defective
60
II
602
Air flow too hot
Air temperature at nasal can-
nula higher than 43°C (external
Measurement mandatory)
60
II
603
ext. Measuring ele-
ment not detected
External measurement element
at nasal cannula not detected,
please check.
60
I
651
Applicator heating
defective
Applicator heating defective.
60
I
652
Applicator not
detected
Applicator not detected, please
change.
60
I
653
Type of applicator!
The selected real flow rate is
too high for this type of
applicator. Reduce flowrate or
use a larger applicator.
70
II
701
System failure
EEPROM damaged, please
restart system. In the case of
failure contact your responsible
TNI medical AG representative.