45
Operating Instruction
TNI
soft
Flow
50
homecare system
Error category
Priority
Error
number
Indication in the
display
Description
20
I
255
Dew point not
reachable
Selected dew point not reach-
able.
(Difference about 1°C, depend-
ing on the set dew point)
30
II
301
Heating plate
overheated
Heating plate overheated.
Please check humidifier.
Detection: Heating plate tem-
perature is hotter than it should
be (Limit depending on set
temperature)
30
I
351
Please refill water
Water chamber empty, fill up
with water.
The water level is not measured
directly, but can be determined
with the heating plate
characteristics.
30
I
352
Heating plate
defective
Heating plate defective (no
power), contact your responsi-
ble TNI medical AG representa-
tive.
30
I
353
Sensor defective
Temperature sensor humidifier
defective
30
I
354
Heating plate
defective
Heating plate defective (no cur-
rent), contact your responsible
TNI medical AG representative.
30
I
355
Sensor defective
System failure, please restart
system. In the case of failure
contact your responsible TNI
medical AG representative
40
II
401
blower defective
Blower blocked, please switch
system off/on. In the case of
failure contact your responsible
TNI medical AG representative
40
II
402
Blower overheated Blower overheated.
Motor temperature is higher
than 120 °C.