![TNI softFlow 50 Скачать руководство пользователя страница 43](http://html.mh-extra.com/html/tni/softflow-50/softflow-50_operating-instruction_1122047043.webp)
43
Operating Instruction
TNI
soft
Flow
50
homecare system
Filter class of dust filter:
G4 (EN 779: 2003)
Dust filter changing cycle:
Please note 3.1.8
Expected operating time
(Expected Service Life) of
TNI
soft
Flow
50
:
3-6 years, depends on daily usage
Attached Oxygen source
Type:
Only medical approved oxygen sources
may be connected (that includes, but
is not limited to IEC 60601-1:2005 and
for homecare use IEC 60601-1-11:2010
compliance). For more information, please
consult the oxygen user manual or your
oxygen retailer.
For handling and adjustment, please refer
to the according user manual.
Max. pressure allowed at the oxygen intake
200 mbar
8) Error codes
SAFETY NOTES
•
Some errors lead to the result that the device can no longer be operated.
•
In the case of a fault contact your responsible TNI medical AG representative.
•
Alarm signals will sound to indicate danger for the user to an error that have occurred.
•
The delay between error condition and error signal can be up to one minute max.
•
The alarm signals are divided into two steps:
o Alarms with low priority (severity code I):
•
1 audible signal sounds,
•
This procedure is repeated cyclically,
•
The error code will be indicated in the display,
•
When the error is cleared the device is again ready to operate,
•
The visual indication in the display cannot be switched off.
o Alarms with medium priority (severity code II):
•
3 audible signals sounds,
•
This procedure is repeated cyclically,
•
The error code will be indicated in the display,
•
These alarms cannot be switched off via menue,
•
The device can no longer be operated,
•
The user is not able to correct the fault,
•
Contact your responsible TNI medical AG representative.
Содержание softFlow 50
Страница 1: ......
Страница 2: ...TNI System ...
Страница 15: ...15 Operating Instruction TNIsoftFlow 50 homecare system 3 5 6 7 1 2 4 9 10 11 12 8 max min ...
Страница 55: ...55 Operating Instruction TNIsoftFlow 50 homecare system Notes ...