SAFETY NOTE
•
Oxygen supports combustion process. Smoking and open fire are not permitted when oxygen
is used during therapy.
•
Several deaths occurred in hospitals in the past due to patients who smoked despite being
treated with oxygen. In order to take a drag, the oxygen tube was removed from the face and
put on the bed. This made it possible for the oxygen to spread across the beddings and the
patient’s clothes. The patient fell asleep and the lit cigarette set the bedding on fire. It was not
possible to fight the fire due to the spread oxygen. It kept burning even after the attempts to
extinguish it. The patient finally died of the burnings.
Connecting oxygen supply
•
The connection to the oxygen source is located on the left side of the device.
•
Connect the
TNI
soft
Flow
50
with the oxygen supply via the oxygen tube supplied.
•
Do not supply oxygen before the
TNI
soft
Flow
50
has started.
•
The required rate of oxygen will be set / adjusted at the oxygen supply.
3.1.8) Replacing dust filter
•
Please change the “dust filter” at regular intervals (at least every 3 month).
•
Take the “dust filter” cover out of the holder (at the backside).
•
A light downward pressure to the flap releases the lock.
•
The dust filter cover can be removed downwards. Remove the “dust filter”.
•
Insert a new dust filter in case of a change.
•
Insert the “dust filter” cover from below and lock it with light pressure onto the upper part.
•
Protect the filter from direct sunlight and humidity.