Operating Instruction
TNI
soft
Flow
50
homecare system
•
The system goes into operation and supplies the required flow rate at the applicator, the air
flow is perceptible, the value is obvious in the display.
•
Supply the
TNI
soft
Flow
50
with the required oxygen, adjust the value.
•
The oxygen applied is shown in the display in l/min.
•
The system mixes the adjusted air-flow rate with the oxygen.
•
The flow rate is referred to as total flow; i.e. the air-oxygen mixture is set to the flow rate
specified in the nominal value.
•
The real value and the oxygen concentration, resulting from it, FiO
2
in %, is shown in the
display.
3.9) Stopping oxygen supply
SAFETY NOTE
•
The connection to the oxygen source needs to be set up correctly as described in chapter
3.1.7.
•
The oxygen supply may only be released with the
TNI
soft
Flow
50
running.
•
Do not supply the
TNI
soft
Flow
50
with oxygen during standby-mode!
•
Please note chapter 3.1.7.
•
The system still supplies the required flow rate at the applicator, the air flow is perceptible.
•
The real flow rate and the FiO
2
, resulting from it, is shown in the display.
•
Switch off the oxygen supply at the oxygen source.
•
The oxygen shown in the display rises to 0 l/min.
•
The FiO2 value shown in the display indicates 21%.
•
The system switches off.
•
The
TNI
soft
Flow
50
now indicates the standby display.