![TNI softFlow 50 Скачать руководство пользователя страница 19](http://html.mh-extra.com/html/tni/softflow-50/softflow-50_operating-instruction_1122047019.webp)
19
Operating Instruction
TNI
soft
Flow
50
homecare system
3.1.5) Inserting the applicator
•
Choose the right applicator (see chapter 3.5). Remove it from
the packaging and insert the applicator into the designated
opening of the device.
Insert the applicator plug from above into the opening and
press down with less force until it snaps in.
•
The locking lever moves to the left stop.
WARNING
In order to avoid burnings,
•
do not modify the applicator in any way
•
make sure that the applicator is not heated up to more than
room temperature (e.g. by means of a blanket, electric fire...)
since this may cause serious injury.
•
do not use insulating sleeves or accessories that have not been
authorized or recommended by TNI medical AG.
SAFETY NOTE
•
In order to minimize interference with the supervised signal, position the heated applicator
tube away from any electronic monitoring electrode (EEG, ECG, EMG, etc.).
3.1.6) Removing the applicator
WARNING
In order to prevent mechanical destruction,
•
remove the applicator from the retainer without effort.
•
do not use force if it is not possible to remove the applicator right away.
•
do not use any tools to remove the applicator.
SAFETY NOTE
•
Do not let any foreign substances or objects get into the system’s
applicator opening.
Removing the applicator
•
Move the locking lever under the applicator to the right.
•
The applicator is released from its locking.
•
Remove the applicator plug upwards from the holder.
•
Now, the system can be used with another applicator.
3.1.7) Connecting oxygen supply
WARNING
In order to prevent burnings,
•
do not modify the applicator in any way.
•
note the safety guidelines in chapter 1.3.6.
•
do not place the
TNI
soft
Flow
50
device on the floor. Keep a minimum distance of 40 cm!
•
keep the minimum distance of 40 cm to the wall.
•
keep the minimum distance of 80 cm to other electric devices.
•
note that an external oxygen source not being completely connected may lead to an inadequate
therapy.
•
Be very careful when handling oxygen! Fire hazard!