6 Hygiene
6.2.4.1 Pre-cleaning
. Disconnect the device from the mains power in standby
mode as described in these operating instructions.
· Unplug the connecting hose from the connecting unit to
the humidifier and the applicator from the humidifier to the
patient.
· Wipe the surfaces of the humidifier unit as well as the
connecting unit with a lint-free cloth moistened with cleaning
solvent.
· Open the humidifier chamber by taking off the housing lid
and removing the humidifier cover as well as the reservoir.
Remove the water from the reservoir.
· Pre-clean the humidifier cover, the reservoir and the
separator panels under running water, immerse them in
a water bath containing bio-degradable cleaning solvent
and wipe with a soft cloth moistened in cleaning solvent.