16
Operating Instructions TNI
®
20s oxy Version 2.7
Version 1.6.3 July 2010
4.1 Assembling and setting up the TNI
®
device
-
Place the humidifier unit near to your bed/place of treatment,
so that it is standing horizontally and is easily accessible.
- TNI
®
20s: Turn the rotary knob of the setting valve at the air
connection
(“AIR inscription“) clockwise (OFF direction) until it
stops. Do not turn the knob any further.
4 Startup
- Interconnect the units as described in the
operating instructions. Do not connect the
units to any other units.
- Do not place the unit near sources of heat
and do not expose it to sunlight! Heat ex-
posure can impair proper functioning.
- Place the humidifier unit horizontally on a
flat and level surface.
- Prior to initial startup, especially check
whether there is any cracking on the
pressure dome. Do not operate if damaged,
hazard of breakage!
- Observe the following safety instructions,
especially for an admixture of clinical oxy-
gen:
· Do not place the humidifier unit on the
floor. Maintain a minimum distance of
40 cm from the floor.
· Do not place the humidifier unit on
any soft surfaces (e.g. foam pad).
· Maintain a minimum distance of 80
cm to other electrical devices.
· Maintain a minimum distance of 40
cm to the wall.
4 Startup
IMPORTANT!