6
Operating Instructions TNI
®
20s oxy Version 2.7
Version 1.6.3 July 2010
3 Safety
3.1
Safety instructions for handling oxygen
- The applicator may not be placed on the TNI
®
unit or any
electrically powered devices.
- Make sure that the seal of the humidifier basin is undamaged
and has been correctly inserted.
- The TNI
®
device may only be operated with the delivered gas
connection units. It is not permitted to operate the apparatus
with other gas connection units. The maximum operating
pressure is 5 bar/500 kPa. The measuring scale for the flow
rates is designed for a pressure of 5 bar and the specified
deviations are maintained under these conditions.
-
Do not let O
2
fixtures come into contact with oil, grease or
combustible liquids. No smoking, no open fires - fire ha
-
zard!
- The device may not be operated in closed areas, where
anesthetic mixtures and oxygen and/or nitrous oxide gas
are manufactured or used.
- Observe the following safety instructions, especially for an
admixture of clinical oxygen:
· Do not place the humidifier unit on the floor. Maintain a
minimum distance of 40 cm from the floor.
· Do not place the humidifier unit on any soft surfaces
3 Safty
Please read the following instructions carefully!
You will find some important hints and tips on
the safe and secure handling and operation of
the TNI
®
device. Special training for initial startup
of the device is not necessary. The device is
operated by healthcare professionals.
IMPORTANT!
It is the user‘s responsibility to observe the
following notices.
Failure to observe these
notices may result in damage or injury!
WARNING!