18
Operating Instructions TNI
®
20s oxy Version 2.7
Version 1.6.3 July 2010
→
→
→
→
→
Humidifier unit with housing lid
Display
Sealed keypad with
keys and rotary
knob
Standby
button
Connection for
applikator
Connection for
connecting hose
Humidifier unit
TNI
®
20s oxy /
back view
Master
switch
US
B connection
(for service only)
4 Startup
installed in combination with the air connection!
- Join the connection fittings of the
connection unit and the
connection of the
humidifier unit with the connecting hose. Avoid
kinks and creases.
- Note that the product variant
TNI
®
20s must not be operated with
oxygen!
A respective warning text
(“AIR ONLY“) is located on the back
of the TNI
®
20s.
Socket
for
sensor cable
Device plug
→
→
→
→
→
- Connect the sensor cable to the provided socket on the back
of the humidifier unit. Do not connect any other components to
this socket.
- First plug the power cable into the device plug on the back of
the humidifier unit and then into the power socket.