8 Error Codes
Some errors and malfunctions mean that the
unit can no longer be operated! If this occurs,
immediately contact the Service Center.
If an error could mean danger for the user or for the patient, an alarm
signal sounds. There are two types of alarm signals:
−
One signal tone
indicates a lower priority error. The error code
is indicated on the display. Once the error is corrected (see
following table) the device is operational.
−
Three signal tones
indicate a high priority. The error code is
indicated on the display and the device can not be used further.
The error cannot be remedied by the user. Please contact the
Service Center immediately.
The following error codes may appear on the display in the event
of malfunctions:
IMPORTANT!
8 Error Codes
Error
code
Cause/results
Remedy
110
Internal error
Contact the Service Center.
111
Internal error
Contact the Service Center.
112
The error may have
the following causes:
No water in the res-
ervoir.
Not enough water in
the reservoir. Water
used up during op-
eration.
Switch TNI
®
20 oxy to standby
mode. Allow the unit to cool down.
Fill the reservoir with water up to
the mark.
Switch TNI
®
20 oxy to standby
mode. Allow the unit to cool down.
Fill the reservoir with water up to
the mark.
113
Internal error
Contact the Service Center.