WSA5000 Functional Overview
Figure 4:
IQ Offset Correction
Table 3:
RF Front-End Control/Status Commands
SCPI Command
Description
:INPut
:ATTenuator[?]
Enables/disables the front-end's 20dB attenuation (on some models)
:VARiable[?]
Sets the variable attenuation for WSA5000-418 and -427
:FILTer
:PRESelect[?]
Enables/disables the use of preselect filtering
:GAIN
:IF[?]
Sets the variable IF gain
:HDR[?]
Sets gain level for the NB ADC of of the HDR signal path
:MODE[?]
Selects the receiver mode of operation
:SOURce
:REFerence
:PLL[?]
Selects the 10MHz reference clock source
:RESET
Resets the 10MHz reference selection to INTernal source
[:SENSe]
:DECimation[?]
Sets the decimation rate as an exponent of 2 (i.e. rate = 2
level
where level =
0,
1,
2 - 10)
:FREQuency
:CENTer[?]
Sets the center frequency of the RFE
:IF?
Queries the IF frequencies that are used for the current input mode and
center frequency
:LOSCillator?
Gets the frequency to be set for the external LO 1 or 2 in corresponding to
current the WSA's center frequency
:RESolution?
Gets the Analog PLL tuning resolution
:SHIFt[?]
Sets the frequency shift value (not available for HDR mode)
:LOCK
:REFerence?
Queries the lock status of the PLL reference clock
:RF?
Queries the lock status of the RFE's PLL
:OUTput
:IQ
:MODE[?]
Selects the IQ output path to be from the external connector or the
digitizer
:CONNector
:INVersion?
Queries if a spectral inversion is required at a given frequency
onward) for further details on each set of
commands.
Digital Signal Processing
The WSA5000 has embedded DSP blocks to provide further signal processing
capabilities, such as DDC with up to 10 levels of decimation and FFT computation.
ThinkRF WSA5000 Wireless Signal Analyzer Programmer's Guide
20