SCPI Command Set
INPut Commands
:INPut:ATTenuator
This command enables, disables or queries the WSA5000's RFE 20dB attenuation.
Note:
This command applies to WSA5000-2XX, -3XX, -408 and their variants only. It is
not applicable for -408P, -418 and -427 and their variants, see
Syntax
:INPut:ATTenuator <Boolean>
:INPut:ATTenuator?
Parameter
ON | OFF | 1 | 0
Input Data Type
Integer | Character
Query Response
1 | 0
Output Data Type
Integer
*RST State
1
Examples
:INP:ATT ON
:INPUT:ATT?
:INPut:ATTenuator:VARiable
This command sets or queries the variable attenuation of the WSA5000's RFE.
Note:
This command applies to WSA5000-408P, -418, -427, and their variants only. For
WSA5000 -2XX, -3XX, -408 and their variants, see
Syntax
:INPut:ATTenuator:VARiable <integer [dB]>
:INPut:ATTenuator?
Parameter
<Attenuation [dB]>
Input Data Type
Integer, optional character unit
Allowable Values
0 – 25 dB
Query Response
<integer>
*RST State
0
Examples
:INP:ATT:VAR 5
:INPUT:ATT:VAR?
This command is not available for sweep mode. The desired attenuation should be set
before sweep mode starts.
:INPut:FILTer:PRESelect
This command sets or queries the RFE preselect filter selection.
60
ThinkRF WSA5000 Wireless Signal Analyzer Programmer's Guide