WSA5000 Functional Overview
SCPI Command
Description
:DWELl[?]
Sets the maximum amount of time waited for a trigger to occur after which
the trigger is aborted
:PPBlock[?]
Same as
:SPPacket[?]
As defined in
:TRIGger
:TYPE[?]
As defined in
:LEVel[?]
As defined in
:PERiodic[?]
As defined in
) for further details.
Synchronized Sweep
The WSA5000 supports a synchronized sweep function for the purposes of comparing
the same signal received via multiple WSA5000s.
Synchronized sweep is an extension of the external trigger capability. One of the
WSA5000s in a network is configured to be the master (
and the other WSA5000s are configured as slaves (
master and slaves are configured with a sweep list, in which each sweep entry has a
synchronization trigger type (
synchronization trigger is generated and delivered from the master's GPIO to that of the
slaves to indicate the beginning of a capture.
provides a synchronization trigger example using sync-word. The master sends
the sync-word when the setup of its front-end has been completed. Master and slaves
are also individually configured with a delay variable (
a resolution of 8 nsec). This delay wait time accounts for the typical worst-case front-end
setup time and for differences in the synchronization cable length. Master and slaves
then begin the capture upon the expiration of the wait (or delay).
ThinkRF WSA5000 Wireless Signal Analyzer Programmer's Guide
26