SCPI Command Set
Syntax
:INPut:MODE ZIF |
DD | HDR |
|
SH | SHN
:INPut:MODE?
Parameter/Response
ZIF | DD | HDR |
| SH | SHN
I/O Data Type
Character
*RST State
Product version dependent
Examples
:INP:MODE HDR
:INPUT:MODE?
1
IQIN mode is not available in WSA5000 product version 2.2. See
to find out your product
version (or the Administrative web-console to the box).
2
SHN mode is only available in WSA5000 hardware revision 3 (product version 1.3) onward. See
to find out your hardware version (or the Administrative web-console to the box).
3
The RFE mode availability is product dependent. Hence, the
state and the initial power-up
default would be different depending on the product version.
4
The HIF mode is only available for WSA5000-XXX-HIF model, which is indicated through the
command
with the response code 002. When this model is used, the
INPut:MODE[?] command will only take/response with the DD and HIF parameters.
SOURce Commands
:SOURce:REFerence:PLL
This command selects and queries the 10MHz reference clock source, whether it be via
the internal source or through the external SMA connector.
Caution:
When the external 10MHz reference is used, its reference level
must
be
between
-10dBm
and
0dBm
.
Exceeding the level of 0dBm will result in permanent
damage to the internal clock circuit. Additionally, the 10MHz reference must be powered
down prior to powering down the WSA5000.
Syntax
:SOURce:REFerence:PLL INT | EXT
:SOURce:REFerence:PLL?
Parameter/Response
INT | EXT
I/O Data Type
Character
*RST State
INT
Examples
:SOURCE:REF:PLL INT
:SOUR:REF:PLL?
:SOURce:REFerence:PLL:RESET
This command resets the 10MHz reference clock in to the internal source.
Syntax
:SOURce:REFerence:PLL:RESET
Parameter/Response
None
*RST State
INT
Examples
:SOURCE:REF:PLL:RESET
ThinkRF WSA5000 Wireless Signal Analyzer Programmer's Guide
63