SCPI Command Set
:SYSTem:OPTions?
This command queries the hardware option(s) or features that a particular WSA model
supported. The response string contains comma separated 3-digit values to represent
the options. See
for the translated list. Examples:
•
WSA5000-2XX-ELO would have code 001 returned in the response string to
indicate “external local oscillators” option;
•
WSA5000-4XX-WBIQ-ELO would have code 001,003 returned in the response
string to indicate ELO and wideband IQ options.
Syntax
:SYSTem:OPTions?
Parameter
None
Response
<xxx>{,<xxx>}
Output Data Type
Comma separated 3-digit value (ex: 000, 001, 002)
*RST State
None
Example
:SYST:OPT?
Table 36:
WSA Option Codes and the Corresponding Description
Option Code
Description
Related SCPI Command
000
No Special Option
001
External Local Oscillator (ELO) Input
[:SENSe]:FREQuency:LOSCillator?
002
WSA5000-XXX-HIF Model
003
WSA5000-4XX-WBIQ
None
004
WSA5000-4XX-P Model
None
1
ELO and/or WBIQ
are special WSA5000 variants, not available on all WSA5000s. Contact
ThinkRF for more details on the usage of these variants.
:SYSTem:SYNC:MASTer
This command sets the WSA unit to be the master or slave for a synchronization trigger
system with multiple units, in which
only one unit
can be the master.
The master sends a sync-word or pulse (set through
or
) via its GPIO to that of the slaves to indicate the
beginning of a capture. The master WSA itself will has an internal loop-back of the
synchronization signal it sent out.
Syntax
:SYSTem:SYNC:MASTer <Boolean>
:SYSTem:SYNC:MASTer?
Parameter
ON | OFF | 1 | 0
Input Data Type
Integer | Character
Query Response
0 | 1
Output Data Type
Integer
*RST State
0
52
ThinkRF WSA5000 Wireless Signal Analyzer Programmer's Guide