imPULSE S.U.M. with Touchscreen Console User’s Guide
|
62
Chapter 3
|
Touchscreen Console setup
3. When you are finished, touch
Enter
. The Pump Module
Configuration screen will reappear with the name you entered next
to the
Pump Name
field with the added prefix (i.e. P1: Acid).
Calibration
Set up a new pump calibration using four-point calibration, and either
mass or volume units.
1. If you would like to calibrate the pump, touch the
Calibration
field.
2. Follow the calibration instructions that are detailed in
Chapter 4—Calibration Procedures.
3. Touch
Done
to return to the Home screen.
3.4.8 Configuring the Fill module
The Mass module must be configured and pumps must be set up prior
to configuring the Fill module. Use the following information and steps
to configure the Fill module.
To begin, locate the Fill module on the Touchscreen Console Home
screen, and touch
Configure
. The Fill module Configuration screen will
appear (Figure 3.34).
Note:
You can touch
Done
at the bottom of the Configuration screen
at any time to save your current settings and so you can resume later.
Figure 3.32. Renaming the Pump 1 module.
Figure 3.33. Renamed
Pump 1 module.