Thermo Scientific
imPULSE S.U.M. with Touchscreen Console User’s Guide
|
29
• 2,000 and 5,000 L units do not require a summing box for load
cells. When making connections, each of the three load cells
can be plugged directly into the Touchscreen Console.
2. Use the auxiliary power ports on the back of the Touchscreen
Console as needed.
Note:
Auxiliary power is at the same
line voltage as the Touchscreen Console, either 120 or 240 V,
depending on your system. Verify that the auxiliary breaker on the
back of the Touchscreen Console is in the “on” position before use.
CAUTION:
All four auxiliary power ports share the same breaker
(8 amps (A) of power out of the Touchscreen Console). To avoid
tripping the auxiliary breaker, do not use higher than 8 A power for all
four of the auxiliary power ports combined.
When you are ready to operate the system, turn on the main breaker,
followed by the main power switch. After the Touchscreen Console has
booted up, verify that the E-Stop is disengaged, then press the blue
“power-on reset” button (Figure 2.30).
Figure 2.30. Pressing the blue
power-on reset button.
Chapter 2
|
Hardware installation