imPULSE S.U.M. with Touchscreen Console User’s Guide
|
57
Chapter 3
|
Touchscreen Console setup
Figure 3.27. BPC Pressure High High and Low Low
Alarms screen.
1. If you would like to set High High/Low Low alarms for air pressure,
touch the
High High and Low Low alarms
field.
2. On the screen that appears, there are separate sections for both
High High and Low Low alarms, each with respective Interlock
and Hysteresis fields (Figure 3.27). To set either alarm, move the
corresponding slider to the
On
position.
3. Touch the
Interlock
field of an enabled alarm to enter the Interlock
equation screen.
4. Fill out the Interlock equation by entering the
Control setpoint
(in either PSI or mbar) and selecting the
Action taken
(either
Pause
or
Stop
).
– If
Pause
is set as the action taken, you must fill out the Hysteresis
equation by entering the
Control setpoint
(in either PSI or mbar)
and the
Time to hold
before operations can resume (in seconds).
– If
Stop
is set as the action taken, operations must be resumed
manually.
5. Touch
Done
to return to the previous screen.
6. Once you finish setting the High High and/or Low Low alarms,
touch
Done
.
7. The system will confirm the settings—touch
Done
to return to the
BPC Module Configuration screen.