imPULSE S.U.M. with Touchscreen Console User’s Guide
|
46
Chapter 3
|
Touchscreen Console setup
Figure 3.16. Mass High and Low alarms screen.
Required fields
There are no required fields for the Mass module.
Optional fields
High and Low alarms
Choose to enable or disable High and/or Low alarms. These alarms
provide a visual alert on the display when conditions are outside of
specified operating parameters.
1. If you would like to set High/Low alarms for mass, touch the
High
and Low alarms
field. The screen that appears has separate
sections for both High and Low alarms (Figure 3.16).
2. To enable a High or Low alarm, move the corresponding slider to the
On
position.
3. Touch the
High
or
Low alarm
field and enter the mass value (in kg)
in the adjacent field that will trigger the High or Low alarm.
4. Touch
Enter
to confirm and return to the previous screen.
5. After you finish setting the High and Low alarm values, touch
Done
.
6. After a moment, a confirmation screen will appear. Touch
Done
to
return to the Configuration screen—the values in the
High and Low
alarms
field will reflect your changes.