![ThermoFisher Scientific thermo scientific imPULSE S.U.M. Скачать руководство пользователя страница 29](http://html1.mh-extra.com/html/thermofisher-scientific/thermo-scientific-impulse-s-u-m/thermo-scientific-impulse-s-u-m_user-manual_1105607029.webp)
Thermo Scientific
imPULSE S.U.M. with Touchscreen Console User’s Guide
|
24
Before making connections
1. Verify that the main breaker and main power switch are in the “off”
position. The selection switch should be pointing to the small circle
to the left (Figure 2.13).
2. Verify that the emergency stop, referred to as the “E-Stop,” is
disengaged (pulled out) by turning the knob clockwise (Figure 2.14).
Note:
The E-Stop disconnects all power to the system. An alarm
buzzer will sound when the E-Stop is activated.
3. Ensure that the E-Stop jumper (SV5142.904—not included) is
plugged into the E-Stop port on the back of the Touchscreen
Console (Figure 2.15).
Figure 2.13. Main power switch
shown in “off” position.
Figure 2.14. Turning knob on the E-Stop
clockwise to disengage.
Chapter 2
|
Hardware installation
Figure 2.15. E-stop Jumper connected
to the E-Stop port.
E-Stop Jumper