Thermo Scientific
imPULSE S.U.M. with Touchscreen Console User’s Guide
|
52
Chapter 3
|
Touchscreen Console setup
2. To set either alarm, move the corresponding slider to the
On
position.
3. Touch the
High
or
Low alarm
field and enter the value (in PSI) in the
that will trigger the High or Low alarm.
4. Touch
Enter
to confirm and return to the previous screen.
5. After you have entered the High and Low alarm values, touch
Done
.
6. After a moment, a confirmation screen will appear. Touch
Done
to
return to the Configuration screen—the values in the High and Low
alarms field will reflect your changes.
High High and Low Low alarms
Choose to enable or disable High High and/or Low Low alarms. These
alarms provide a visual and audible alert, and will pause or stop the
specified pumps based on the set Interlock equations.
1. If you would like to set High High/Low Low alarms for air pressure in,
touch the
High High and Low Low alarms
field.
2. On the screen that appears, there are separate sections for both
High High and Low Low alarms, each with respective
Interlock
and
Hysteresis
fields (Figure 3.23). To set either alarm, move the
corresponding slider to the
On
position.
Figure 3.22. BPC Pressure High and Low Alarms screen.