imPULSE S.U.M. with Touchscreen Console User’s Guide
|
58
Chapter 3
|
Touchscreen Console setup
Figure 3.28. BPC Alarm Designation (Pumps) screen.
Alarm designation (pumps)
Use this field to specify which pump(s) are affected by the High High
and/or Low Low alarms.
1. If you would like to designate alarms for pumps that have been
set up, touch the
Alarm designation
(pumps)
field on the BPC
Module Configuration screen.
Note:
Enabling a pump allows it to adjust the air pressure after the
alarm has been triggered.
2. On the screen that appears (Figure 3.28), you will see the
interlock and hysteresis parameters that you entered for the High
High and/or Low Low alarm(s). Below that, the pumps that have
been set up with the system are shown. Touch a
pump
option.
3. On the screen that appears, move the corresponding slider to the
On
position for the High High and/or Low Low alarms to turn on the
alarm(s) for the selected pump.
4. When you are finished, touch
Done
. You will be returned to the
BPC Alarm Designation screen.
5. After setting all desired alarm designations touch
Done
.
6. After a moment, a confirmation screen will appear. Touch
Done
to
return to the BPC Module Configuration screen—the values in the
Alarm designation field will reflect your changes.