Survey Scan/Scanning
The Survey Scan program allows you to scan a
wavelength range. When you use the Survey Scan
program, you can perform these tasks:
•
Recall a test
OR
set up your test parameters
•
Run the Cell Correction program
•
Collect a baseline scan
•
Scan an unknown
•
View scan data
•
Change the scale of the data plot
•
Determine 3-point net measurements
•
Calculate the area under a curve
•
Label peaks and valleys
Note:
The Survey Scan/Scanning program allows
you to measure only one unknown at a
time. Auto 6, Auto 3 and Manual 6 are not
available for scanned measurements.
Note:
If you want to set a baseline expiration
time, press
UTILITIES
, then use the arrow
keys to highlight
Baseline Expiration
.
Press
ENTER
and set the desired time.
To get started, press the
TEST
key on the keypad.
When the
Test Types
screen appears, highlight
Survey Scan
or
Scanning
and press
ENTER
.
Recalling a test
1.
With the
Survey Scan/Scanning
screen
displayed, press
Stored Tests
. A list of stored
tests appears.
2.
Use the arrow keys to highlight the name of
the test you want to recall and press
ENTER
.
The parameters for the selected test appear on
the screen.
From this screen, you can:
•
Set up test parameters
•
Set up Cell Correction program
•
Save a test
•
View the list of stored tests
•
Collect a baseline
•
Scan unknowns
Setting up test parameters
1.
With the
Survey Scan/Scanning
screen
displayed, use the arrow keys to highlight the
name of the parameter you want to set.
2.
When the parameters are set, you can press
Save Test
to save the test or
Run Test
to
measure a blank or an unknown.
Collecting a baseline scan
Note:
If your instrument is equipped with a
6-Position Cell Holder, be sure to place the
blank in the B position. The instrument always
uses the B position to collect the baseline.
1.
With the
Survey Scan/Scanning
test
parameters displayed, press
Run Test
.
2.
Place the blank in the B position.
3.
Press
Measure Blank
to collect the baseline.
When the instrument is finished measuring the
blank, the message disappears.
Note:
If you want to switch between tabular and
graphical displays, press
Graph/Tabular
.
Using “General Tests” Software
Thermo Electron Corporation
3-20
Содержание 335904
Страница 1: ...BioMate 3 Series Spectrophotometers Operator s Manual ...
Страница 2: ......
Страница 8: ......
Страница 12: ...Thermo Electron Corporation iv ...
Страница 60: ...Using General Tests Software Thermo Electron Corporation 3 28 ...
Страница 70: ...Thermo Electron Corporation 5 4 Connecting Using Accessories ...
Страница 74: ...Performing Maintenance Procedures Thermo Electron Corporation 6 4 ...
Страница 80: ...Appendix B Parameters Thermo Electron Corporation B 4 ...
Страница 88: ...Appendix C Calculations Thermo Electron Corporation C 8 ...
Страница 92: ...Index Thermo Electron Corporation I 4 ...
Страница 93: ......
Страница 94: ...Item No 335904 10001 Rev G 06 03 Component No 335904 10063 Rev G 06 03 ...