Note:
If
Cell Correction
is
ON,
you must run the
Setup Correction
program before you can
access the
Run Test
or
Measure
Samples
keys.
Setting up test parameters
1.
With the
Direct UV (280)
setup screen
displayed, use the arrow keys to highlight the
name of the parameter you want to set.
2.
When the parameters are set, you can press
Save Test
to save the test or to measure the
blank and unknowns.
Measuring the sample
Measuring unknowns automatically (using Auto 6
or Auto 3)
1.
With the
Direct UV (280)
screen displayed,
place the blank and unknowns in the correct
cell positions.
2.
Press
ENTER
to start the measurements. The
instrument automatically measures the blank
first, then measures the unknowns and
displays the absorbance and concentration of
each unknown.
Measuring unknowns manually (using Manual 6 or
Single Cell Positioner)
1.
With the
Direct UV (280)
screen displayed,
press
Run Test
.
2.
Place the blank and unknown in the correct cell
positions. If your instrument is equipped with a
6-Position Cell Holder, you can place up to five
samples in the cell holder.
3.
Press
Measure Blank
to measure the blank. If
your instrument is equipped with a 6-Position
Cell Holder, it automatically moves to the B
position to measure the blank. When it
completes the measurement, it returns to its
previous cell position.
4.
Press
Measure Sample
to start the
measurement. If your instrument is equipped
with a 6-Position Cell Holder, press the cell
position buttons on the keypad to re-position
the cell holder and measure the rest of the
unknowns manually.
Warburg-Christian
The Warburg-Christian analysis uses an
absorbance difference measurement at 280 and
260nm to determine the protein concentration of an
unknown. Refer to Appendix B for a description of
the parameters and Appendix C for the default
values.
To get started, move the arrow keys on the
BioMate tests
screen to highlight
Protein Tests
and press
ENTER
. Then the move the arrow keys
to highlight
Warburg-Christian
and press
ENTER
.
The
Warburg-Christian
parameter screen
appears.
Note:
If
Cell Correction
is
ON
, you must run the
Setup Correction
program before you can
access the
Run Test
or
Measure
Samples
keys.
Setting up test parameters
1. With the
Warburg-Christian
setup screen
displayed, use the arrow keys to highlight the
name of the parameter you want to set.
2. When the parameters are set, you can press
Save Test
to save the test or
Run Test
to
measure the blank and unknowns.
Using “Biotests” Software
Thermo Electron Corporation
2-10
Содержание 335904
Страница 1: ...BioMate 3 Series Spectrophotometers Operator s Manual ...
Страница 2: ......
Страница 8: ......
Страница 12: ...Thermo Electron Corporation iv ...
Страница 60: ...Using General Tests Software Thermo Electron Corporation 3 28 ...
Страница 70: ...Thermo Electron Corporation 5 4 Connecting Using Accessories ...
Страница 74: ...Performing Maintenance Procedures Thermo Electron Corporation 6 4 ...
Страница 80: ...Appendix B Parameters Thermo Electron Corporation B 4 ...
Страница 88: ...Appendix C Calculations Thermo Electron Corporation C 8 ...
Страница 92: ...Index Thermo Electron Corporation I 4 ...
Страница 93: ......
Страница 94: ...Item No 335904 10001 Rev G 06 03 Component No 335904 10063 Rev G 06 03 ...