Measuring the sample
Measuring unknowns automatically (using Manual
6 or Single Cell Positioner)
1.
With
the Warburg-Christian
screen displayed,
place the blank and unknowns in the correct
cell positions.
2.
Press
ENTER
to start the measurements. The
instrument automatically measures the blank
first, then measures the unknowns and
displays the absorbance and concentration of
each unknown.
Measuring unknowns manually (using Manual 6 or
Single Cell Positioner)
1.
With the
Warburg-Christian
screen displayed,
press Run Test.
2.
Place the blank and unknown in the correct cell
positions. If your instrument is equipped with a
6-Position Cell Holder, you can place up to five
samples in the cell holder.
3.
Press
Measure Blank
to measure the blank. If
your instrument is equipped with a 6-Position
Cell Holder, it automatically moves to the B
position to measure the blank. When it
completes the measurement, it returns to its
previous cell position.
4.
Press
Measure Sample
to start the
measurement. If your instrument is equipped
with a 6-Position Cell Holder, press the cell
position buttons on the keypad to re-position
the cell holder and measure the rest of the
unknowns manually.
5.
When the instrument has measured all the
unknowns, it displays the absorbance and
concentration.
Cell Growth
The cell growth measurement uses absorbance at
600nm to indicate the progress of cell growth in a
sample. The instrument does not perform any
calculations or graphing for the data.
To get started, move the arrow keys on the
BioMate Tests
screen to highlight
Cell Growth
and press
ENTER
. The
Cell Growth
setup screen
appears.
Note:
If
Cell Correction
is
ON
, you must run the
Setup Correction
program before you can
access the
Run Test
or
Measure
Samples
keys.
Setting up test parameters
1.
With the
Cell Growth
setup screen displayed,
use the arrow keys to highlight the name of the
parameter you want to set.
2.
When the parameters are set, you can press
Save Test
to save the test or
Run Test
to
measure the blank and unknowns.
Measuring the sample
Measuring unknowns automatically (using Auto 6
or Auto 3)
1.
With the
Cell Growth
measurement screen
displayed, place the blank and unknowns in
the correct cell positions.
2.
Press
ENTER
to start the measurements. The
instrument automatically measures the blank
first, then measures the unknowns and
displays the absorbance of each unknown
Measuring unknowns manually (using Manual 6 or
Single Cell Positioner
)
1.
With the
Cell Growth
measurement screen
displayed, press
Run Test
.
2.
Place the blank and unknown in the correct cell
positions. If your instrument is equipped with a
6-Position Cell Holder, you can place up to five
samples in the cell holder.
3.
Press
Measure Blank
to measure the blank. If
your instrument is equipped with a 6-Position
Cell Holder, it automatically moves to the B
position to measure the blank. When it
Using “Biotests” Software
2-11
BioMate 3 Operator’s Manual
Содержание 335904
Страница 1: ...BioMate 3 Series Spectrophotometers Operator s Manual ...
Страница 2: ......
Страница 8: ......
Страница 12: ...Thermo Electron Corporation iv ...
Страница 60: ...Using General Tests Software Thermo Electron Corporation 3 28 ...
Страница 70: ...Thermo Electron Corporation 5 4 Connecting Using Accessories ...
Страница 74: ...Performing Maintenance Procedures Thermo Electron Corporation 6 4 ...
Страница 80: ...Appendix B Parameters Thermo Electron Corporation B 4 ...
Страница 88: ...Appendix C Calculations Thermo Electron Corporation C 8 ...
Страница 92: ...Index Thermo Electron Corporation I 4 ...
Страница 93: ......
Страница 94: ...Item No 335904 10001 Rev G 06 03 Component No 335904 10063 Rev G 06 03 ...