
To edit the concentration of a standard
1.
With the standard curve displayed on your
screen, use the arrow keys to highlight the
standard you want to edit. Press
Edit
Standards
.
2.
With the
Edit Concentration
highlighted,
Press
ENTER
.
3.
Press
Edit Conc
or a number key.
4.
Enter the concentration value in the
Entry
field.
5.
When the concentration value is correct, press
ENTER
to accept the value.
To add a standard
1.
With the standard curve displayed on your
screen, press
Edit Standards
.
2.
Use the arrow keys to highlight
Add Standard
.
3.
Enter the concentration value of the additional
standard in the
Entry
field.
4.
When the concentration value is correct, press
ENTER
to accept the value.
5.
Press
Measure Standards
to remeasure all
the standards.
To delete a standard
1.
With the standard curve displayed on your
screen, use the arrow keys to highlight the
standard you want to delete. Press
Edit
Standards
.
2.
Use the arrow keys to highlight
Delete
Standard
. Press
ENTER
to delete the
standard.
To clear measurements
1.
With the standard curve displayed on your
screen, press
Edit Standards
.
2.
Use the arrow keys to highlight
Clear
Measurements
and press
ENTER
. All the
absorbance measurements will be removed
from the screen.
To reset standards
1.
With the standard curve displayed on your
screen, press
Reset Standards
and press
ENTER
.
2.
Use the arrow keys to highlight
Reset
Standards
and press
ENTER
. All the
standards and measurements will be removed.
To select a different curve fit for a standard curve
Note:
To change the curve fit for a standard
curve, you must display the standard curve
as a graph, not as a table.
1.
With the standard curve you want to edit
displayed as a graph on your screen, press
Change Fit
.
2.
Use the arrow keys to highlight the curve fit
you want to use for the standard curve and
press
ENTER
. The instrument applies the
selected curve fit to the data and displays the
new graph.
Absorbance Ratio
When you use the Absorbance Ratio program, you
can perform these tasks:
•
Recall a test
OR
set up your test parameters
•
Run the Cell Correction program
•
Measure a blank
•
Measure unknowns
To get started, press the
TEST
key on the keypad.
When the
Test Types
screen appears, highlight
Absorbance Ratio
and press
ENTER
.
Using “General Tests” Software
Thermo Electron Corporation
3-14
Содержание 335904
Страница 1: ...BioMate 3 Series Spectrophotometers Operator s Manual ...
Страница 2: ......
Страница 8: ......
Страница 12: ...Thermo Electron Corporation iv ...
Страница 60: ...Using General Tests Software Thermo Electron Corporation 3 28 ...
Страница 70: ...Thermo Electron Corporation 5 4 Connecting Using Accessories ...
Страница 74: ...Performing Maintenance Procedures Thermo Electron Corporation 6 4 ...
Страница 80: ...Appendix B Parameters Thermo Electron Corporation B 4 ...
Страница 88: ...Appendix C Calculations Thermo Electron Corporation C 8 ...
Страница 92: ...Index Thermo Electron Corporation I 4 ...
Страница 93: ......
Страница 94: ...Item No 335904 10001 Rev G 06 03 Component No 335904 10063 Rev G 06 03 ...