When you use a factor, you need to specify the
factor and the concentration units.
•
When you use a standard, you need to specify
the concentration of the standard and measure
its absorbance.
When you use the Basic Concentration program,
you can perform these tasks:
•
Set the wavelength
•
Measure a blank
•
Measure a standard
OR
enter a factor
•
Measure unknowns
The steps for taking measurements in the two
modes are similar - the only difference will be
whether you measure a standard or enter a factor.
Setting the wavelength & mode
1.
Press
Set nm
or any number key to set the
wavelength.
2.
Enter the wavelength where you want the
measurements taken, then press Set nm again.
3.
Press
Change Mode
until the appropriate
measurement mode (Concentration with
Standard or Concentration with Factor)
appears.
Measuring a blank
1.
Place the blank in the cell holder. If your
instrument has a 6-Position Cell Holder, be
sure to place the blank in the B position.
2.
If you want to enter a concentration for the
blank, press a number key and enter the
concentration in the
Entry
field.
3.
Press
Measure Blank
to measure the blank.
When the instrument is finished measuring the
absorbance of the blank, the message
disappears.
Measuring a standard
1.
If necessary, press
Change Mode
to switch to
the
Concentration with Standard
mode.
2.
If your instrument is equipped with the
6-Position Cell Holder, place the standard in
one of the cell positions and press the
corresponding cell position key on the keypad.
If your instrument is equipped with the Single
Cell Holder Platform, remove the blank and
place the standard in the cell holder.
3.
Press
Units/Standard
to set the units and
measure the standard.
4. Enter Conc
, use the number keys to enter the
concentration value of the standard and then
press
ENTER
.
5.
Press
Select Units
, use the arrow keys to
highlight the entry in the
Units
field and then
press
ENTER
to select the units for the
concentration.
6.
Press
Measure Standard
to measure the
standard. When the instrument is finished
measuring the absorbance of the standard, it
displays the absorbance and calculated factor.
Using “General Tests” Software
3-7
BioMate 3 Operator’s Manual
Содержание 335904
Страница 1: ...BioMate 3 Series Spectrophotometers Operator s Manual ...
Страница 2: ......
Страница 8: ......
Страница 12: ...Thermo Electron Corporation iv ...
Страница 60: ...Using General Tests Software Thermo Electron Corporation 3 28 ...
Страница 70: ...Thermo Electron Corporation 5 4 Connecting Using Accessories ...
Страница 74: ...Performing Maintenance Procedures Thermo Electron Corporation 6 4 ...
Страница 80: ...Appendix B Parameters Thermo Electron Corporation B 4 ...
Страница 88: ...Appendix C Calculations Thermo Electron Corporation C 8 ...
Страница 92: ...Index Thermo Electron Corporation I 4 ...
Страница 93: ......
Страница 94: ...Item No 335904 10001 Rev G 06 03 Component No 335904 10063 Rev G 06 03 ...