Entering a factor
1.
If necessary, press
Change Mode
to switch to
the
Concentration with Factor
mode.
2.
Press
Units/Factor
to set the factor and select
the units.
3.
Use the arrow keys to highlight the entry in the
Factor
field and press
Enter Factor
to enter
the factor.
4.
Use the number keys to enter the factor.
5.
Press
Enter Factor
to accept the factor and
return to the screen displaying the factor and
units. The factor you just entered appears on
the display.
6.
Press
ESC
to return to the
Concentration
with Factor
screen.
Measuring unknowns
1.
If your instrument is equipped with the
6-Position Cell Holder, place the unknown you
want to measure in one of the cell positions
and press the corresponding cell position
button on the keypad to move the cell holder to
the measuring position. The measurement
appears on the display.
If your instrument is equipped with the Single
Cell Holder Platform, remove the blank and
place the unknown in the cell holder. The
measurement appears on the display.
Advanced A/%T/C - Absorbance &
%Transmittance measurements
When you use the Advanced A/%T/C program for
absorbance or %transmittance measurements, you
can perform these tasks:
•
Select the measurement mode you want to use
(Absorbance or %Transmittance)
•
Run Cell Correction program
•
Recall a test OR set up your test parameters
•
Measure a blank
•
Measure unknowns
To get started, press the
TEST
key on the keypad.
When the
Test Types
screen appears, highlight
Advanced A/%T/C
and press
ENTER
.
Recalling a test
1.
With the
Advanced A/%T/C
screen displayed,
press
Stored Tests
. A list of stored tests
appears.
2.
Use the arrow keys to highlight the name of
the test you want to recall and press
ENTER
.
The parameters for the selected test appear on
the screen.
Using “General Tests” Software
Thermo Electron Corporation
3-8
Содержание 335904
Страница 1: ...BioMate 3 Series Spectrophotometers Operator s Manual ...
Страница 2: ......
Страница 8: ......
Страница 12: ...Thermo Electron Corporation iv ...
Страница 60: ...Using General Tests Software Thermo Electron Corporation 3 28 ...
Страница 70: ...Thermo Electron Corporation 5 4 Connecting Using Accessories ...
Страница 74: ...Performing Maintenance Procedures Thermo Electron Corporation 6 4 ...
Страница 80: ...Appendix B Parameters Thermo Electron Corporation B 4 ...
Страница 88: ...Appendix C Calculations Thermo Electron Corporation C 8 ...
Страница 92: ...Index Thermo Electron Corporation I 4 ...
Страница 93: ......
Страница 94: ...Item No 335904 10001 Rev G 06 03 Component No 335904 10063 Rev G 06 03 ...