Note:
When Cell Correction is activated,
additional parameter lines are added to the
screen above the
Cell Correction
line. If
the
Cell Correction
line is no longer
visible on the screen, highlight
More
parameters...
and press
ENTER
.
3.
Highlight
Set Up Correction
and press
ENTER
. The
Cell Correction
screen appears.
4.
Highlight
Correction Mode
and press
ENTER
to set the mode to either:
•
Scan
- Cell Correction is run on a blank and
one sample cell for the range of
wavelengths you specify in
Scanning
mode.
•
Discrete nms
- The Cell Correction
program is run on a blank and up to five
sample cells for up to 31 user-specified,
discrete wavelengths.
5.
If you selected Scan mode in the preceding
step, press
Run Corr.
to start the Cell
Correction program. If you selected Discrete
nms mode, first specify the wavelengths using
the procedures which follow, and then the Cell
Correction program.
The Cell Correction program will measure the other
cells against the blank and will record, store and
date the measurements. From these
measurements the Cell Correction program
establishes the required correction factors, which
then are automatically applied during all
subsequent tests (if Cell Correction is activated).
Specifying wavelengths for Discrete nms mode:
1.
Highlight
Sample Positioner
and press
ENTER
to set this parameter to either
Auto 3
(when using three, large cell holders) or
Auto
6
(when using six, small cell holders).
2.
Highlight
Number of Matched Cuvettes
and
press
ENTER
. Then use the keypad to specify
the number of cells you are matching. Press
ENTER
.
3.
Press
Set nms
to select the wavelengths for
which the Cell Correction program will be run. A
list of wavelengths appears.
Using “General Tests” Software
Thermo Electron Corporation
3-4
Содержание 335904
Страница 1: ...BioMate 3 Series Spectrophotometers Operator s Manual ...
Страница 2: ......
Страница 8: ......
Страница 12: ...Thermo Electron Corporation iv ...
Страница 60: ...Using General Tests Software Thermo Electron Corporation 3 28 ...
Страница 70: ...Thermo Electron Corporation 5 4 Connecting Using Accessories ...
Страница 74: ...Performing Maintenance Procedures Thermo Electron Corporation 6 4 ...
Страница 80: ...Appendix B Parameters Thermo Electron Corporation B 4 ...
Страница 88: ...Appendix C Calculations Thermo Electron Corporation C 8 ...
Страница 92: ...Index Thermo Electron Corporation I 4 ...
Страница 93: ......
Страница 94: ...Item No 335904 10001 Rev G 06 03 Component No 335904 10063 Rev G 06 03 ...