Chapter 8
Maintenance & Troubleshooting
4. Holding the pyrolyzer tube in place, apply the Kaowool insulation over
the top of the insulation.
Figure 8-17. Apply the Kaowool insulation
5. Holding the insulation and pyrolyzer tube, gently attach the housing lid
against the insulation, allowing both ends of the tube to come out of
the lid approximately 1/2”. Attach one of the graphite ferrules and one
of the fittings removed earlier to each of the pyrolyzer tube ends.
Figure 8-18. Reattach the ferrules and fittings
6. Gently slide the lid back into the housing and secure it using the screws
and washers removed earlier.
7. With an air gun placed at the end of one fitting, blow out any dust or
debris that may have dropped inside the pyrolyzer quartz or glass tube
(
8. Follow steps 22 through 27 in the previous section,
, to re-install the housing and complete this procedure.
Thermo Scientific
SOLA II Flare User
Manual
8-17
Содержание SOLA II Flare
Страница 1: ...SOLA II Flare Sulfur Online Analyzer User Manual PN 90 1312 0...
Страница 2: ......
Страница 3: ...SOLA II Flare Sulfur Online Analyzer User Manual PN 90 1312 0 Revision A...
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 14: ......
Страница 18: ......
Страница 42: ...Chapter 3 Start Up Shutdown Thermo Scientific SOLA II Flare User Manual 3 6 Figure 3 4 Diagnostics Menu...
Страница 44: ......
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 80: ......
Страница 104: ......
Страница 122: ......
Страница 136: ......
Страница 148: ......
Страница 158: ......
Страница 160: ......
Страница 162: ......