Chapter 5
Configuration
7. The Purge alarm will remain active for a default time of 15 minutes at
startup or after the purge is restored.
A l a r m s
S e t u p
P u r g e
A l a r m
T i m e o u t
∨
∧
E n t
E x i t
Figure 5-23.
8. Set the Status Alarms Mode to Latching or Non Latching. This value
allows the user the option to latch an alarm condition, or not, to allow
for after-the-fact troubleshooting of intermittent problems or user
acknowledgment of an alarm.
A l a r m s
S e t u p
S t a t u s
A l a r m s
M o d e
∨
∧
E n t
E x i t
Figure 5-24.
Calibration
Overview
The analyzer requires a two-point linear calibration. During the calibration
procedure the average of the raw detector signals corresponding to two known
sulfur concentrations are latched by the unit and retained in non-volatile
memory.
During normal operation, the average detector reading is interpolated on the
line to derive the sulfur concentration reported.
Thermo Scientific
SOLA II Flare User
Manual
5-8
Содержание SOLA II Flare
Страница 1: ...SOLA II Flare Sulfur Online Analyzer User Manual PN 90 1312 0...
Страница 2: ......
Страница 3: ...SOLA II Flare Sulfur Online Analyzer User Manual PN 90 1312 0 Revision A...
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 14: ......
Страница 18: ......
Страница 42: ...Chapter 3 Start Up Shutdown Thermo Scientific SOLA II Flare User Manual 3 6 Figure 3 4 Diagnostics Menu...
Страница 44: ......
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 80: ......
Страница 104: ......
Страница 122: ......
Страница 136: ......
Страница 148: ......
Страница 158: ......
Страница 160: ......
Страница 162: ......