Appendix F
The X-Purge System
AC Power
AC power to the X-Purge is connected by the customer. Power wiring and
circuit breakers must be sized appropriately. Refer to drawings provided with
the system for connection information and power requirements.
Customers must provide a suitable power switch near the system for use by
maintenance personnel.
Warning
Electrical power must be free of spikes, sags, surges or electrical
noise.
▲
AC power to any system using the X-Purge is connected directly to the X-Purge
unit rather than the analyzer. The X-Purge unit controls the power to the
instrument to ensure safe operation in hazardous areas. Consult the following
table for ac power connections to the X-Purge unit.
Table F–4.
Power
Terminal
Hot
TB1-1
1
Neutral
TB1-3
2
Ground
TB1-5
Power out (switched)
TB1-6
1
Terminals 1 and 2 are jumpered together.
2
Terminals 3 and 4 are jumpered together.
Thermo Scientific
SOLA II Flare User
Manual
F-4
Содержание SOLA II Flare
Страница 1: ...SOLA II Flare Sulfur Online Analyzer User Manual PN 90 1312 0...
Страница 2: ......
Страница 3: ...SOLA II Flare Sulfur Online Analyzer User Manual PN 90 1312 0 Revision A...
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 14: ......
Страница 18: ......
Страница 42: ...Chapter 3 Start Up Shutdown Thermo Scientific SOLA II Flare User Manual 3 6 Figure 3 4 Diagnostics Menu...
Страница 44: ......
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 80: ......
Страница 104: ......
Страница 122: ......
Страница 136: ......
Страница 148: ......
Страница 158: ......
Страница 160: ......
Страница 162: ......