Appendix A
Watlow EZ Zone PM6 Series
Table A–18. Oven settings: Setup Page > Control Loop 1
Parameter
Default
Comment
Display
Software
Display
Software
Heat Algorithm
PID
The method of heat control.
Cool Algorithm
Off
TRU-TUNE+ Enable
No
TRU-TUNE+ Band
0°C
TRU-TUNE+ Gain
3
Autotune Aggressiveness
Critical
Critical damped aggressiveness of
autotuning calculations.
Peltier Delay
0.0
User Failure Action
User
When user switches to manual mode,
controller sets output power to last
open-loop set point entered by user.
Input Error Failure
Off
When an input error switches control
to manual mode, controller sets
output to 0%.
Fixed Power
0.0%
Open Loop Detect Enable
No
Open Loop Detect Time
240
Open Loop Detect Deviation
6.0°C
Ramp Action
Off
Ramp Scale
Minutes
Ramp Rate
1.0°C
Low Set Point
0°C
The minimum value of the closed loop
set point range.
High Set Point
130.0°C
The maximum value of the closed loop
set point range.
Set Point Open Limit Low
-100.0%
The minimum value of the open loop
set point range.
Set Point Open Limit High
100.0%
The maximum value of the open loop
set point range.
Thermo Scientific
SOLA II Flare User
Manual
A-11
Содержание SOLA II Flare
Страница 1: ...SOLA II Flare Sulfur Online Analyzer User Manual PN 90 1312 0...
Страница 2: ......
Страница 3: ...SOLA II Flare Sulfur Online Analyzer User Manual PN 90 1312 0 Revision A...
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 14: ......
Страница 18: ......
Страница 42: ...Chapter 3 Start Up Shutdown Thermo Scientific SOLA II Flare User Manual 3 6 Figure 3 4 Diagnostics Menu...
Страница 44: ......
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 80: ......
Страница 104: ......
Страница 122: ......
Страница 136: ......
Страница 148: ......
Страница 158: ......
Страница 160: ......
Страница 162: ......